The range function in the LCR driver is not working. The issue is that we attempt to perform a fancy string to number to string coversion. The LCR will only take the following in its SCPI commands (tested manually)
📖 SCPI Command Reference
The driver wraps these low-level serial commands into high-level Python methods.
| Command | Parameter Options | Description | Example Usage | Example Return (Blank means no return value) |
|---|---|---|---|---|
*IDN? |
- | Identity check | *IDN? |
Agilent Technologies,U1732C,... |
SYST:ERR? |
- | Error queue check | SYST:ERR? |
+0, "No error" |
MODE |
SER, PAL |
Series vs Parallel | MODE SER |
- |
DISP2 |
D, Q, TH |
Set secondary parameter | DISP2 Q |
- |
FREQ |
100, 120, 1k, 10k |
Test frequency (Hz) | FREQ 1k |
- |
FUNC |
R, C, L, Z, ESR, AI |
Primary measurement type | FUNC C |
- |
RANG |
See Range Table below | Set manual range | RANG 20u |
- |
FETC? |
- | Read primary value | FETC? |
+6.856E-07 |
FETC? ALL |
- | Read all parameters | FETC? ALL |
Rs,+3.809052E-13,Ls,+4.316664E+09,Rp,+6.172276E-14,Lp,+4.316802E+09,Z,+4.316733E+09,TH,+5.774222E-01,F,+1.000000E+03,D,+1.774207E+02,Q,+5.636322E-03 |
📉 Range Options by Function
| Function | Available Range Arguments |
|---|---|
| R / Z | 2, 20, 200, 2k, 20k, 200k, 2M, 20M, 200M |
| L | 2000u, 20m, 200m, 2, 20, 200, 2k |
| C | 2000p, 20n, 200n, 2000n, 20u, 200u, 20m |